next previous
Up: The high energy

2. Instrument description

 

A sketch of the major components of the PDS experiment is shown in Fig. 1 (click here), and a summary of its main features is given in Table 1 (click here).

  table236
Table 1: Main features of the PDS experiment

The PDS consists of a square array of four independent NaI(Tl)/CsI(Na) scintillation detectors. A system of anticoincidence (AC) shields surrounds the sides and the front of the detectors and collimators, that limit the telescope field of view (FOV). There are two independent collimators, one for each pair of detectors. Each collimator bench can be independently rocked back and forth to allow the simultaneous monitoring of source and background. The gain of the detectors and of the associated electronic chain is continuously monitored and automatically controlled by using as sensor a fixed calibration source (FCS) of tex2html_wrap_inline1324. The instrument can also be periodically in-flight calibrated with a movable calibration source (MCS). The AC lateral shields are also used as gamma-ray burst monitor. Finally a particle monitor (PM) provides information on high environmental particle fluxes and triggers actions to prevent the damage of the detectors photomultipliers (PMT). All events detected in the PDS detector are processed by an Analog Processor (AP), that includes all the analog electronics, analog-to-digital converters (ADCs), AC logics, programmable thresholds. Each detected event qualified as a good one by the AP is time tagged and sent to a Digital Processor (DP) for the data processing via a 40 bit FIFO. The DP filters and formats the data that are finally sent via Block Transfer Bus to the BeppoSAX On-Board Data Handling (OBDH). The AP and DP constitute the PDS Instrument Controller (IC).

In the following subsections the main elements of PDS are described.

2.1. Detection plane

The detection plane of PDS is an array of four scintillation detectors. Each of the four detectors is composed by two crystals of NaI(Tl) and CsI(Na) optically coupled and forming what is known as PHOSWICH (acronym of PHOsphor sandWICH). The NaI(Tl) thickness is 3 mm and the CsI(Na) thickness is 50 mm.

The scintillation light produced in each phoswich (PHW) is viewed, through a quartz light guide, by a PMT EMI D-611. The NaI(Tl) acts as the X-ray detector, while the CsI(Na) scintillator acts as an active shield. The scintillation events (due to photons or charged particles) detected by each phoswich can be divided into three classes: those that deposit energy only in the NaI crystal (NaI events or ``good events''); those that deposit energy only in the CsI crystal (CsI events); those that deposit energy in both crystals (mixed events). The events of the first class are accepted, while the events of the other two classes are rejected. The rejection is performed using the different decay constants of the scintillations produced in the NaI (about tex2html_wrap_inline1326 at room temperature) and CsI (tex2html_wrap_inline1328 and about tex2html_wrap_inline1330 for CsI). The analysis electronics (see Sect. 2.7 (click here)), via its Pulse Shape Analyzer (PSA), is capable to discriminate events of the classes described above. Figure 2 (click here) shows a typical pulse shape spectrum in the 15-300 keV range. It is characterized by two peaks, one (on the left) due to NaI(Tl) events and the other due to CsI(Na) events. The scale in abscissa is correlated with the decay time of the scintillation pulses.

  figure255
Figure 2: Typical pulse shape spectrum in the 15-300 keV energy range. Note the discrimination between NaI ``good'' events (left) and CsI events (right)

The selection of good events is performed through the choice of lower and upper programmable PSA thresholds in the Electronic Unit (see below).

The phoswich configuration is well known to provide a high detection efficiency and a very low background level. In addition, taking into account that the accepted photons lose energy in the NaI(Tl) alone, the energy spectra provided by phoswiches are not affected by biases as in single crystal detectors. The total geometric detection area is 795 cmtex2html_wrap_inline1336. The other crystal parameters are given in Table 1 (click here).

2.2. Collimator assembly

The collimator assembly consists of two X-ray collimators made of Tantalum tubes with hexagonal section. Inside each tube, a 4 cm bi-layer of Tin plus Copper is inserted. The role of Tin, tex2html_wrap_inline1340 thick, is to attenuate K-fluorescence emission from the Tantalum (tex2html_wrap_inline1344 keV, tex2html_wrap_inline1346 keV), while the Copper, tex2html_wrap_inline1348 thick, was used to attenuate the K-fluorescence emission from Tin (tex2html_wrap_inline1352 keV, tex2html_wrap_inline1354 keV). This cell configuration is known as graded shielding. The detector net area through collimators is 80% of the total geometric area.

The collimators limit the detector field of view to 13 Full Width at Half Maximum (FWHM). Each of the two collimators can be independently rocked back and forth with respect to the neutral position directed along the detector axis. Five symmetrical rocking steps are provided on each side of the neutral position. The maximum rocking angle is tex2html_wrap_inline1356 (3.5 degrees) and the intermediate steps are tex2html_wrap_inline1358, tex2html_wrap_inline1360, tex2html_wrap_inline1362, tex2html_wrap_inline1364.

The standard observation strategy is to continuously monitor background and source+background with both collimators using a cyclic law: while one of the two collimators points to the source the other one points to a blank (non source-contaminated) sky field for background measurement. At each cycle the two collimators are swapped: the one pointing to the source is moved to monitor background and viceversa. The default collimator law samples alternatively two background fields on the two opposite sides of the observed source at the maximum rocking angle. Therefore the default full cycle is ON /+210'(OFF) / ON /-210' (OFF). As the collimator law is fully programmable, any other convenient choice can be made. As an example, if one of the two background fields is contaminated by the presence of an unwanted source, a suitable law that rocks the collimators only on one side, or that changes the rocking angle to a value which excludes contamination, can be chosen. It is also possible to move the collimators through all the allowed positions. In this way one can derive the angular response of the collimators to a given source and the source position can be determined.

The dwell time (minimum 50 s) in each collimator position can be chosen from ground.

A position transducer gives an independent high accuracy (about 1 arcsec) monitoring of the collimator position for each collimator. This monitoring is transmitted to ground in the housekeeping (HK) data stream with a sampling rate of 1 s.

2.3. Detector window

The X-rays entering through the telescope FOV cross the top shield materials and the entrance window of the phoswich crystals (kept under vacuum). The top shield window is made of tex2html_wrap_inline1370 BC-434 plastic scintillator covered with aluminized Kapton (tex2html_wrap_inline1372 Al plus tex2html_wrap_inline1374 Kapton). The phoswich window is made of tex2html_wrap_inline1376 Be, 890 tex2html_wrap_inline1378m Silicone rubber, aluminized Kapton (tex2html_wrap_inline1380 Al plus tex2html_wrap_inline1382 Kapton) and tex2html_wrap_inline1384 Teflon film. The Silicone rubber was used to provide thermal and mechanical insulation, the Teflon film was used as light reflective material, while the aluminized Kapton was used as a barrier between the Silicone rubber and the reflective material.

2.4. Anticoincidence shield system

A further reduction in the intrinsic background of PDS is achieved using a five sides anticoincidence shield system. This consists of the lateral and top shield assemblies, both highly efficient in the rejection of events due to charged particles. The CsI(Na) lateral shields are also very efficient as active shields of photons coming from the sides.

In the following we describe this subsystem and the anticoincidence logics used to operate it.

2.4.1. Lateral shield assembly

The lateral shielding system is made of four optically independent slabs of CsI(Na) scintillators 10 mm thick, 275 mm high and 402 mm wide. Each slab is made of two pieces optically coupled together. Each slab is viewed, through a quartz light pipe, by two 3'' PMTs (Hamamatsu R-2238) with independent high voltage power supplies. In order to monitor the gain of each detector, a light pulser made of an tex2html_wrap_inline1388 source encapsulated in a small NaI(Tl) scintillator is inserted into each slab to produce pulses equivalent to photons with energy within the range from 350 to 450 keV, depending on the particular scintillator slab. The summed signals from each slab are sent to a programmable threshold for the AC logic. If the summed signal is above the AC threshold, a veto signal is generated by the PDS analog processor.

2.4.2. Top shield assembly

The top shield assembly consists of a 1 mm thick plastic scintillator (BC-434) viewed by four PMTs (Hamamatsu R-1840) with independent power supplies. It is located above the collimator assembly and covers the entire telescope FOV. This shield was included in the PDS design to reject the background events due to charged particles entering the aperture of the instrument that deposit part of their energy in the top shield. In particular we expect that this shield is effective in rejecting events generated by high energy electrons. These electrons are expected to be produced by interactions of Cosmic Rays with the satellite materials.

The summed signal from the four PMTs is sent to a programmable analog threshold for the AC logic. If the summed signal is above the AC threshold, a veto signal is generated by the PDS analog processor.

2.4.3. Anticoincidence shield with other PHW units

Any event that is detected from one of the four PHW units is compared with a programmable analog AC threshold. If the signal is above this threshold, a logic signal is set. A veto signal is generated by the PDS analog processor if at least two units set this logic signal. This happens when two or more simultaneous events are detected from the PHW units above their A/C thresholds. In this way we reject events that interact with one of the NaI scintillators via Compton effect, leaving only a fraction of their energy in it, while the photon with the remaining energy is absorbed in other PHW unit(s).

2.4.4. AC logic and operative modes

The AC system acts as a logical OR among the AC veto signals issued by the PDS Analog Processor. The decision time of the AC logic is programmable with 5 equispaced steps between 2 and tex2html_wrap_inline1390 independently for each of the AC signals described in the previous sections. The AC logics are permanently activated with two operative modes selectable from the ground: AC-ON in which the vetoed events are not transmitted to ground; AC-OFF, in which the vetoed events are flagged and transmitted to ground. This last mode can be used only when PDS is transmitting data in some Direct Modes (see Sect. 2.7 (click here)). The AC of each PHW unit, of the lateral shields and of the front shields can be separately programmed to be ON or OFF.

2.5. Automatic gain control

The automatic gain control of the phoswich detectors (AGC) is implemented as follows. At the center of the detection plane there is a fixed calibration source (FCS). It consists of an tex2html_wrap_inline1392 radioactive source distributed within a plastic scintillator (BC-400) viewed by a PMT (Hamamatsu R 1635). Part of the 60 keV X-rays emitted by this source impinges on the four phoswich detectors. These events are tagged thanks to the 5.5 MeV tex2html_wrap_inline1394 particles simultaneously emitted with them and detected by the plastic scintillator.

The rate of 60 keV photons detected by each phoswich as pure NaI events ranges from 3.6 to tex2html_wrap_inline1396 depending on the phoswich unit.

The signal from the PMT coupled to the plastic scintillator is used by the PDS electronics to tag the photons simultaneously detected by the PHWs. The tagging efficiency is about 98%. The tagged photons are used by the PDS IC to detect gain variations and to perform a continuous and automatic gain correction by acting on the high voltage supplies of the phoswich PMTs. This correction is operated using two prefixed contiguous channel windows, with the separation channel corresponding to the tex2html_wrap_inline1398 60 keV peak position at nominal gain. Each time a NaI event tagged as tex2html_wrap_inline1400 event is inside the left window or the right window, the HV of the corresponding phoswich is changed by +97 mV or -97 mV, respectively. The spectra of the tagged tex2html_wrap_inline1406 photons are transmitted with the housekeeping information with 128 s integration time.

The AGC is operated by the PDS DP. It is therefore completely reprogrammable from ground, if needed.

The FCS is not suitable to be used as an absolute calibrator, since the FCS photons impinge only on a small part of the phoswich detector surfaces, close to the centre of the detection plane. Due to disuniformity in the scintillation light collection across the crystal surface, a bias in the detector gain should derive. Absolute gain measurement can be performed using the in-flight calibration system described in the next section.

2.6. In-flight calibration

The in-flight calibration of the PDS is performed by means of two different devices: a Movable Calibration System (MCS) and a light pulser.

2.6.1. Movable calibration system

The movable calibrator MCS is made of a radioactive source of tex2html_wrap_inline1408 (main lines at 14.37, 122.06 and 136.47 keV) distributed along a wire of about 32 cm length, and contained in a cylinder of lead that shields the radioactive source but a small aperture directed toward the crystal assembly. The MCS, upon an on-ground command, performs a scan of the entire field of view of the instrument above the collimators and top shield at constant speed, uniformly irradiating all the PHWs. Given the limited half life of the source (271 days), the calibrator crossing speed can be varied in order to get the needed statistical quality of the source data.

2.6.2. Light pulser

A light pulser (LED) is encapsulated in each phoswich crystal. It allows to simulate a prefixed gamma equivalent energy (GEE) lost in the phoswich. In this way the gain of the phoswich PMTs and electronic chain can be checked. The LED is mainly used for diagnostic purposes.

2.7. Electronic unit

 

The electronic unit includes a power supply module and, as discussed above (see Sect. 2 (click here)), the AP, which is devoted to analog data processing and digital conversion, and DP. This has in charge two tasks: the PDS data management and the instrument control. The DP includes two microprocessors (type 80 C86): one (the Event Processor, EP) to perform the first task and the other (the Communication Processor, CP) to perform the second task. Both processors make use of hybrid circuits.

The pulses from each phoswich unit with amplitude in the band corresponding to the operative energy band of PDS that pass the AC logic are pulse height and pulse shape analyzed.

The AC logic allows to reject one or more of the following types of events detected by the phoswich units:

  1. Events simultaneous with events above the AC threshold of the lateral shields;
  2. Events simultaneous with events above the AC threshold of the top shield;
  3. Events simultaneous with events detected by other phoswich units with energy above their AC threshold.

If one or more of the above coincided events are accepted, they can be transmitted with the proper AC flag in case of Direct Transmission Mode. By default the AC logic rejects all the event types from 1 to 3 or any combination of them.

The pulse shape analysis is performed using a constant fraction/crossover PSA. For a detailed description of the PSA and of its performance see Frontera et al. (1993).

The pulse height analysis is performed by a 12 bit successive approximation analog-to-digital converter (ADC) with the sliding scale technique. This technique was chosen to improve the average differential linearity of the pulse height channels. The ADC conversion time is 5 tex2html_wrap_inline1410s.

The selection of the good events, corresponding to energy losses only in the NaI(Tl) scintillator, is performed through the setting of lower and upper pulse shape digital thresholds. These thresholds are operated by the EP.

The qualified events are formatted by the EP in different transmission modes before being acquired by the On-Board Data Handling (OBDH). The transmission modes can be direct and indirect. Direct Modes (DM) transmit information about each qualified event detected by PDS. The information that can be transmitted to ground is:

This maximum information can be compressed in two ways: by masking some bits (the less significant bits) of each field or by eliminating one or more fields. Fourteen predefined compressions are available on board: the maximum event size is 5 bytes (40 bits); the minimum event size is 2 bytes (16 bits).

In Indirect Spectral Mode, the qualified events are accumulated in energy spectra histograms, one for each detection unit. For these integrated spectra the number of energy channels (128 or 256), the maximum content of each channel (8 or 16 bits), and the accumulation time (5 choices, between 1 and 32 s) can be independently set. The indirect spectral modes can be run in parallel with temporal modes. These modes allow the transmission of histograms with count rate time profiles in 1 to 4 energy bands with time resolution down to 1 ms.

In addition to the information associated with good events, other housekeeping (HK) information is provided by the DP and transmitted. This includes pulse shape spectra of the phoswich units in the entire energy band, energy spectra of the FCS, various ratemeters, dead time counters, status information and technological data (high voltage and low voltage checks, temperature of different instrument components, status of the MCS and collimator motors). The HK spectra are summed on time intervals of 128 s, the ratemeters are integrated over 1 s, while the technological data are sampled and transmitted with variable rate, from 16 to 64 s depending on the data type.

2.8. Gamma-ray burst monitor

The lateral shields of the PDS are also used as detectors of gamma-ray bursts of cosmic origin (GRB). The signals from each of the slabs in a programmable energy band, nominally 60-600 keV (the low threshold can be lowered down to about 20 keV), are sent to both a Pulse Height Analyser and a GRBM logic unit, where the trigger condition is checked. A GRB event is triggered when at least two of the four lateral shields meet the trigger condition. In this way we expect to reduce the rate of false triggers caused by interactions of high energy particles in the shields. If a GRB condition is verified, the burst time profiles in the entire energy band from each of the four detectors are stored, together with the absolute time of the trigger occurrence. The integration time for count accumulation will be 7.8 ms for 10 s preceeding the trigger, 0.48 ms for 10 s immediately following the trigger and 7.8 ms from 10 s to about 2 minutes after the trigger time. Along with burst profiles, energy spectra integrated over 128 s and ratemeters integrated over 1 s in the GRBM energy thresholds are continuously acquired.

Parameters of the trigger logic are programmable from the ground. Details on GRBM and its performance will be published in a separate paper in preparation.

2.9. Particle Monitor (PM)

A particle detector, made of BC-434 plastic scintillator viewed by a PMT (R-1840), is mounted close to the PDS instrument position in the satellite. The detector has a cylindric-like shape with a diameter of 2 cm and a thickness of 5 mm. It is covered on the top and sides with an aluminum frame of 2 mm thickness. This configuration provides an energy threshold that for electrons is about 1.2 MeV and for protons is 20 MeV. The PM is used to switch off the high voltage supplies of all the PMTs, excluding the PM PMT itself, when the environmental charged particle count rate integrated over 32 s will exceed a programmable threshold and to switch them on again when this level is below another programmable threshold.


next previous
Up: The high energy

Copyright by the European Southern Observatory (ESO)
web@ed-phys.fr