Table 3 summarizes the main parameters for the final 16
mosaics: for each mosaic are listed the number of fields composing it
(columns
rows), the tangent point (sky position used for geometry
calculations) and the synthesized beam (size and position angle).
The spatial resolution can vary from mosaic to mosaic
depending on the particular array (6A, 6C or 6D) used in the
observations. The mean value for the synthesized beam
is
.
![]() |
Figure 3:
Histogram of the (residual) flux in one of the ![]() |
The last three columns of Table 3 show the results of the
noise analysis.
For each mosaic we report the minimum (negative) flux (
Smin)
recorded on the image (typically
|Smin| is of the order of 0.5 mJy,
corresponding to the value at which we have stopped the cleaning)
and the noise level. This has been evaluated either as the FWHMof the Gaussian fit to the flux distribution of the pixels (in the range
),
in order to check for correlated noise (
), or
as the standard deviation of the average flux in several
source-free sub-regions of
the mosaics,
in order to verify uniformity (
).
As expected, the noise distribution is fairly uniform within
each mosaic
and from mosaic to mosaic (
variations). Also, for each mosaic, the two
noise values are consistent, that is the noise can be considered Gaussian (see also Fig. 3).
On average the noise level is
Jy. The typical detection
limits for the
ATESP survey are thus
mJy at
and
mJy at
.
Dynamic range problems can cause slightly higher noise levels of
Jy around strong sources (
Speak> 50-100 mJy).
Such problems appear to be serious only in one mosaic (fld69to75):
the region around the
bright radio source PMNJ0104-3950 suffers from a very high noise level and
a number of spurious sources are present. This was due to strong phase
instabilities during
the observations which could not be removed by self-calibration.
This region (of size
)
was masked
and therefore excluded from further analysis. Excluding this region,
the total unifom sensitivity area covered by the ATESP survey is 25.9 sq. degr. or
sr.
Another problem we faced was the possible presence of artefacts in the images, like spurious sources ("ghosts'') at a level of 0.1 - 0.5% opposite to bright sources with respect to the phase center of the image and `holes' in the centre of the field. The first problem, caused by the Gibbs phenomenon, arises from the use of an XF correlator and can be serious in high dynamic range continuum observations (like ours). The second effect is a system error produced by the harmonics of the 128 MHz sampler clock at 1408 MHz. Both effects can be completely removed as long as the observing bands are centered appropriately (Killeen 1995; Sault 1995). Unfortunately, at the time of our first two observing runs these effects were not yet known. We therefore could apply the corrections only to the data taken in the last observing run.
We point out that the corrections, when applied, result in a larger bandwidth
smearing effect, since only
MHz channels are used
(instead of
MHz).
Wherever not corrected for, the "ghosts'' problem is
unlikely to be serious in our case, since "ghosts'' appear
in different places for each field and so they tend to average out
when mosaicing the fields.
Moreover, only radio sources
brighter than mJy can produce detectable "ghosts'' in the final
mosaics and such bright sources are very few in the region surveyed (
)
and therefore could be easily checked. No evident "ghosts'' have been
found.
We have also tried to reduce the sampler clock self-interference
effect as far as possible by flagging residual bad visibilities (that is
correlated noise) after a first
step of cleaning and self-calibration (see also Sect. 4.2),
but some
of our fields still show it to a small extent. On the
other hand the area of sky affected by "holes'' is of the order of a few
percents (
)
of the total region observed.
As already mentioned in Sect. 4.2, when the UV coverage is incomplete, the cleaning process can "steal'' flux from real sources and redistribute it on top of noise peaks producing spurious ones.
To quantify the actual effect in our mosaics we performed a set of simulations
by injecting point sources in the survey UV data at random positions.
Then the whole cleaning process was started. The number of sources
injected in a mosaic ()
was chosen in order to reduce the
statistical uncertainty without changing significantly the components/image
statistics. The source fluxes cover the entire range of the survey: 250
sources at 3
,
100 at
,
50 at
,
50 at
,
25 at
,
10 at
,
10 at
,
10 at
.
Taking into account the time and frequency resolution and the baseline lenghts, we get about 1500-2500 indepentent UV points for each field observed. This means that with about 2000 independent (not too close together or on top of each other) clean components we could clean the image to zero flux. The average number of (not independent) clean components per mosaic ranges between 1500 and 3200. Since we used a clean loop gain factor g=0.1, the number of independent clean components per mosaic can be estimated as 1/10 of the numbers reported above. We thus expect a significant clean bias effect (10 - 20%), larger for mosaics with a higher number of cleaning components. We then decided to test three mosaics, with a low, an intermediate and a high average number of cleaning components (cc's) respectively: fld34to40 (1616 cc's), fld44to50 (2377 cc's) and fld69to75 (3119 cc's).
The results of the tests are presented in Figs. 4 and 5
and summarized in Table 4.
Figure 4 shows, for each of the three mosaics,
the average source flux measured after the cleaning
(
Soutput) normalized to the true source flux
(
Sinput) as a function
of the flux itself (expressed in terms of ). In general the clean
bias increases going to fainter fluxes and, as expected, depends on the
number of cleaning components. In the best case (1616 cc's) we get
flux underestimation for the faintest sources; in the worst case (3119 cc's)
the effect rises up to
.
The dependency of the clean bias on the number of cleaning components is more
clearly shown in Fig. 5.
Here we present
Soutput/Sinputas a function of the average number of clean components for different source
fluxes (,
,
,
etc.).
Again, the clean bias affects more seriously the faintest sources.
Moreover it is evident that we are not dealing with a linear effect: a
sudden worsening appears at fluxes of the order of
and when
the number of cleaning components exceeds
.
A first order fit of the clean bias effect for the three different mosaics
has been obtained by applying the least squares method to the function
.
The values obtained for the parameters a
and b are listed in Table 4 and the curves are shown in
Fig. 4.
Copyright The European Southern Observatory (ESO)