The analysis of the point run data undergoes various phases for determining the final experimental values of the anode currents.
The energy flux and the photon current, respectively, at the telescope mirrors before being reflected have been calculated as described above. For these calculations an error of 10% has been deduced for the energy flux above the atmosphere (Torres [1987]), of 2% for the extinction (Carter & Clegg [1994]; Hayes et al. [1975]) and of 1% for the z-angle dependency during the measurement yielding in an error of 10.2%. The theoretical anode currents are then calculated in a straight forward way considering the known characteristics of the different parts of the telescope, i.e. spectral reflectivity of the mirrors and the aluminum cones, spectral quantum efficiency and mean gain of the PMTs. This step of the calculation incorporates an error of 8.4% which is typical for the apparatus used. Considering all uncertainties in the calculation of the theoretical anode current a total error of approximately 13% is determined.
In Fig. 2 (top graph)
the measured mean direct anode currents of
each star
Im is plotted versus the theoretically
expected currents
Ith.
![]() |
Figure 2: Calculated anode currents Ith (top graph) and incident photon current IPh (bottom graph) of the stars given in Table 1 versus the measured dc currents Im. A linear least square fit is plotted for both distributions |
The errors assigned to the measured currents arise from statistics in the grid points and
from the systematic uncertainties due to the background brightness in the surrounding
pixels (see above). In spite of the various assumptions and uncertainties the data points
represent very nicely a linear relationship. Using a linear least squares fit one obtains
For the final goal of energy calibration of IACTs, the response function has to be transferred to pulse operation. The necessary general capability of the electronic circuits is given as pointed out before. For purposes of convenience and tracing the procedure the measured anode current is first related with the calculated energy flux. This is very helpful when selecting stars for angular calibration point runs. A compilation of predicted anode currents for various stellar parameters is given in Karschnick ([1996]).
In the second step
the mirror area and the spectral energy flux have to be
considered to calculate the spectral photon current
IPh . This quantity is plotted in the bottom graph of Fig. 2.
The parameters of the linear correlation are given by
Copyright The European Southern Observatory (ESO)