A coded mask imager was developed for laboratory tests. Here the
prototype detector with the RENA readout, a prototype
coded mask (MURA), and a small source of 41 keV X-rays, 0.75 mm diameter
153Gd, were mounted in an optical bench, and their geometry was
adjusted so the divergent X-ray beam magnified the mask's unit cells to
1300 microns at the detector. Since the shadowgram size is
pixels
and the detector is
pixels, the mask/source combination was moved
to several positions relative to the detector to record the full shadowgram.
The shadow of the mask was cast upon the
detector, and the counts versus position along one strip is
shown in Fig. 3. The transitions from blocked to
unblocked mask elements are at precisely the correct positions, showing
the detector's high spatial accuracy, and the modulation transfer is 100%.
Unblocked counts are consistent with a constant value, which
demonstrates the uniformity of response across the detector.
Shadowgrams of the mask were measured by the detector and then deconvolved to obtain a correlation image. Figure 4 shows the resulting shadowgram and image. All the mask features were accurately recorded by the shadowgram. Systematic noise in the image is undetectably small and limited to <0.5% of the source peak. Thus, systematic errors can be expected to be minimal in the reconstruction and analysis of images. See Matteson et al. (1998b) for further information on imaging results.
Copyright The European Southern Observatory (ESO)