ISOCAM was designed to provide images of the sky and polarization
measurements in the m band. It features two
detectors, one for short wavelengths (SW:
m band), the
other for long wavelengths (LW:
m band). The camera has
two channels which cannot be used simultaneously: a selection wheel
holding Fabry mirrors can direct the light beam from the ISO
telescope toward either one of the detectors. The selection
wheel also holds two internal calibration sources which can illuminate
the detectors quasi-uniformly for flatfield purposes. In order to choose the
observing configuration, there are two wheels for each channel. The
first wheel holds four lenses allowing the choice of the spatial sampling:
1.5, 3, 6, and 12'' per
pixel. At ISOCAM wavelengths, the spatial resolution is diffraction
limited. The second wheel holds a dozen discrete band pass
filters with spectral resolution ranging from 2 to 15,
and continuously variable filters (CVF) with spectral resolution of 45.
The sixth wheel, the entrance wheel, has four positions: one
hole and three polarizers. It is possible to observe data with different
exposure times (0.28, 2.1, 5.04, 6.02, 10.08, 20.16 and 60.2 s) due to
the telemetry flow and on-board electronics. The electronic gain
can be adjusted to 1, 2 or 4. The operating temperature of the camera
is as low as 2.4 K, provided by liquid helium
cooling. All details about ISOCAM, including in flight performances, are
available in Cesarsky et al. (1996).
For one observation, the ISOCAM instrument delivers a set of frame pairs (start of integration, called Reset, and end of integration,
EOI). For the long wavelength detector (LW), the signal corresponds
to a simple difference between EOI and Reset. For the short wavelength
detector (SW), it is more complex, and several operations such as
"cross talk correction'' must be done. We assume that these
corrections have been applied to the data being considered
here. We have a set of data noted D(x,y,t,c): one
measurement per pixel position (x,y),
repeated t times, with c configurations (there is a new
configuration each time the pointing position, the filter, the
integration time, etc, are changed).
In the ideal case, calibration will consist of
In practice, however, we have to take into account several problems:
Studies have been done and continue in order to resolve all of these problems, and some solutions have been proposed.
In this paper we present a selection of methods which are commonly used prior to any scientific analysis of the data. Except for the automatic flat field calculation which requires a raster mode, they can be applied to beamswitching, cvf, raster, or polarization measurements. All these methods have been naturally selected by the CAM Instrument Dedicated Team and CAM Instrument Support Team during the last two years of intensive experimentation and data reduction. Other methods also exist, but were found less efficient, or less reliable than those presented here. They are however described in the different CEA and IPAC technical reports given in the reference list.
In the calibration process proposed here, pixels are considered to be independent. This assumption is not true, due to electronic and photonic effects, and the calibration process should be considered as a first approximation. The main area in which this assumption clearly leads to incorrect results is transient correction and, to a lesser extent, cosmic ray removal.
Copyright The European Southern Observatory (ESO)