As is well known, the transmission of light through a
Fabry-Perot interferometer is given by the Airy function:
where R is the coating reflectance, t the spacing
between the plates, the refractive index of the material
between the plates (air in our case),
the angle of
incidence, and
the wavelength of the incident
light.
For fixed , t and
, the transmitted light
consists of a series of narrow peaks (orders), the separation of
which in wavelength is the Free Spectral Range (FSR), and
for small
:
Let us suppose that the interferometer plates are
perfectly flat and parallel (t constant), and that
the interferometer is illuminated by a perfectly collimated
raybundle, normally incident on the plates ().
In this case, the Full Width at Half Maximum
(FWHM) of each order of the FPI
only depends on the reflectance, and with sufficient accuracy,
for R greater than about 0.85, it is:
Due to the periodicity of its passband, an FPI obviously
demands a prefilter to be used as a spectroscopic device.
In the case of an emission spectrum, with well spaced lines,
an interference filter may be sufficient, but in the case of an
absorption spectrum, like the solar one, a single order of
the interferometer must be isolated. For this reason,
the passband of the order sorter must be narrower than the
interferometer FSR, which in turn, through Eqs. (2) and (3),
depends on the required FWHM of the transmission order.
If the use of the FPI is not limited to the
strongest solar lines, the isolated order must be
significatively narrower than a typical photospheric line,
with a FWHM 100
200 mÅ. If we impose
20 mÅ,
corresponding to a spectral resolution
300000,
a FSR
1 Å is obtained for the interferometer.
An order sorter with a FWHM
1 Å will be therefore
necessary.
As described in Paper I, on the IPM a Universal Birefringent
Filter is used as order sorter for the FPI. Each passband of the UBF
has a shape which resembles a sinc2 function, with a FWHM
= 0.250 Å and a FSR = 128 Å at H.
An interference filter (IF), with a FWHM
50 Å,
is then sufficient to isolate a single UBF passband.
From the spectroscopic point of view, the
instrument therefore consists of three filters mounted in series.
If ,
and
respectively are the IF, UBF and FPI spectral transmissions,
the overall instrumental profile
) is:
When the maximum of the
UBF transmission profile coincides with one of the FPI
orders, it follows that the overall instrumental profile
consists of one main peak and some side lobes.
These unwanted transmission bands contribute to the
parasitic light, defined as:
where and
respectively are the
wavelength of the transmission peak and of the first zero of
the UBF passband.
As shown in Paper I, the parasitic light can be
minimized by a suitable matching of the interferometer
FSR with the spacing of the UBF zeroes.
However, it must be taken into account that,
decreasing the operating wavelength, the
relative wavelength position of the UBF passband zeroes
shrinks with respect to the maxima of the FPI transmission
orders.
For extended operating wavelength ranges, this therefore prevents
the minimization of the parasitic light contribution from
the UBF side lobes, that is obtained by adjusting the adjacent
orders of the FPI on the first zeroes of the UBF passband.
A realistic compromise is obtained by
choosing an interferometer with a 3 mm plate separation.
In this case, the side orders of the FPI coincide with
the third zeroes of the UBF passband at H
(FSR = 0.75 Å). The parasitic light is minimum
(
0.7%) at 6563 Å, and maximum (
2.2%)
at the blue edge of the useful wavelength range (4600 Å).
In practice, however, the aforesaid suppression of the unwanted orders of the interferometer is not secured if the UBF passband sensibly differs from the theoretical one and/or the tuning between the UBF and the FPI is not correct. These conditions therefore require a very careful research of the tune solutions for the nine crystal groups forming the UBF (see Paper I), and a wavelength stability of the UBF and of the FPI such as to secure variations of the parasitic light and of the overall passband transparency smaller than the required photometric precision (1%).
For a relative detuning of 9 mÅ at 5500 Å, the peak
transparency of the instrumental profile decreases of 1%,
while the parasitic light increases of 0.4%.
As the interferometer is very stable in wavelength (see
Sect. 2.3), this result requires a wavelength stability
of the UBF better than about 10 mÅ.
This is obtained by measuring the resistance of a
platinum wire, wound on the crystal groups, and by applying small
corrections to the angular position of each group.
The drifts of the UBF passband can then be compensated,
securing a wavelength stability of 1 mÅ, better
than required.
As shown in Sect. 2.1, to reduce the parasitic light, a
maximum plate separation of 3 mm must be adopted for the
interferometer. This condition imposes a lower limit to the
FWHM of the instrumental profile (Eq. (3)), lowering therefore the
attainable spectral resolution. In particular, for a coating
reflectance R = 0.95 (see Table 1 (click here)):
Although the reflective resolution is much larger
than the demanded one
( 300000), we have to consider that
in practice, if the raybundle associated with
a given image point contains a range of
values and/or covers an area of the FPI plates where,
due to flatness errors, the plate separation
t fluctuates, the transmission profile will be broadened
with respect to the reflective one (see later).
To evaluate the effective spectral resolution, we have
therefore to consider the geometry of the rays incident on
the interferometer plates.
Manufacturer | Queensgate Instruments Ltd. |
Type | Mod. ET50 |
piezo-scanned and | |
capacity servo-controlled | |
Clear aperture | 50 mm |
Plate separation | 3 mm |
Wedge angle | 15' |
Coating | Multilayer broadband |
Wavelength range | 4000 ![]() |
Reflectivity | 0.95 at 5500 Å |
Flatness errors of | ![]() |
each plate at 5461 Å | after coating |
Cavity tuning | 2000 nm |
Resolution | 12 bits |
Settling time | ![]() |
(response to step input) | |
Temperature sensitivity | 3 mÅ ![]() |
of the interferometer | |
Temperature sensitivity | 0.1 mÅ ![]() |
of the CS100 controller | |
Electronic noise equivalent | 10 pm Hz-1/2 rms |
displacement of interferometer | |
plates with CS100 | |
Two different optical mountings can be adopted.
In the case of the so called classical mounting,
the image formed by the telescope is
collimated and the FPI is placed near the image of the
entrance pupil. Each image point is then formed by a
camera lens, which focuses a beam of rays propagating through
the interferometer at the same angle with respect to the optical
axis. In this case, the FWHM of each order
of the FPI is larger than the reflective one
().
In fact, the raybundle associated with each image point
covers an area of the FPI
plates within which the plate separation t fluctuates;
for
= 0, this is accounted for by the
term in the following equation:
where p are the flatness errors (defects) in fractions
of .
Moreover, for what concerns the off-axis image points,
the largest effect is a shift of the transmission orders
towards the blue (),
following the well known equation:
In classical mounting, therefore, the final image plane is
spectrally dishomogeneous, in the sense that the wavelength of
each order is not the same on each point, but
depends on its distance from the optical axis,
and this is also true for a perfect interferometer.
Instead, in the case of the so called
telecentric mounting, the entrance pupil is collimated and the
FPI is placed near an image plane.
Consequently, all the image points are formed by ray cones
normally incident on the interferometer and
containing all the possible directions allowed by the optics.
In this case, the FWHM of each order
of the FPI is larger than the reflective one,
because the raybundle corresponding to each image point
contains a range of values.
Moreover, as the plate separation is not infinitely small,
the ray cones always cover a small, but finite area of the FPI plates,
and therefore, on each point of the final image plane,
the FWHM of each order is also broadened by small scale
flatness fluctuations.
In practice, to avoid that the interferometer plates are focused on the final image, the FPI is not placed on the image plane, but as far away as possible from it. In this case, each ray cone corresponding to each image point covers an area of the interferometer plates, the size of which depends on the distance of the FPI from the image and on the relative aperture of the beam. As the beam incident on the interferometer has a very small aperture (see later), the area covered by each ray cone is only some millimeter in size.
In conclusion, the FWHM () of the
interferometer orders will be broadened by the flatness errors as
well as by the varying
, accounted for by the
and
terms respectively on the following equation:
where are the small scale flatness errors
in fractions of
.
Generally p is defined as twice the maximum deviation of the
considered surface from its ideal mathematical reference
surface; this parameter combines therefore coarse and fine
shape errors.
However, in telecentric mounting, the broadening of the
instrumental profile on each point of the final focal plane
is produced by flatness errors, the scale of which is
significatively smaller than the area of the interferometer
plates. For this reason in Eq. (9) instead of p,
has been used, which, as previously said, just
represents the surface fluctuations at small scale.
Since obviously
, the
term in
is generally smaller than the
corresponding one in
; in other words,
the broadening of the instrumental profile produced by
flatness errors is generally larger in classical than in
telecentric mounting.
Also in the case of telecentric mounting, the final image plane is spectrally dishomogeneous, but in the sense that the wavelength position and the FWHM of each order randomly changes from a point to an other, due to the t fluctuations. However, unlike the case of the classic mounting, this is true only for a real interferometer.
Between the two optical mountings, the telecentric
one has been adopted for the IPM. In this case
it is possible to minimize the spectral dishomogeneities
of the final image plane and to contain the broadening of the
instrumental profile, produced by the varying ,
within the limit imposed by the demanded spectral resolution.
As a matter of fact,
for the adopted interferometer (see Table 1 (click here)),
a surface accuracy of
/150 after coating
at 5461 Å is claimed by the manufacturer.
As a consequence, the
maximum wavelength fluctuation of the instrumental
profile on the final image plane is 13.3 mÅ peak to peak
(728 ms-1) at 5500 Å.
Moreover, assuming
,
, R = 0.95, and imposing a spectral resolution
at 5500 Å, we obtain from Eq.
(9)
, i.e. an f/190 relative aperture of the
beam incident on the interferometer.
In classic mounting, instead, an exceedingly small relative
aperture (f/820) should be used to have the same wavelength
variation from the center to the edge of the field (Eq. (8)).
In practice, an f/192 relative aperture has been adopted,
corresponding to
.
Assuming = 5500 Å,
= 1,
, R = 0.95,
,
(see Table 1 (click here)), we find:
where the inequality symbol is due to the assumption
.
Following Eq. (9), we may conclude therefore that
the FWHM of the instrumental profile and the
spectral resolution of the IPM are:
As the overall instrumental profile essentially is one order of the interferometer, its wavelength stability essentially depends on the interferometer itself, and, in particular, on the stability of the optical length of the cavity, i.e. the product between the geometrical distance of the plates and the refractive index of the air between them.
On the adopted interferometer, the plate separation, as well their parallelism, are stabilized by a capacitance servo-control. A three channel bridge system, the CS100, uses capacitance micrometers to derive error signals when the spacing or the parallelism change. These signals are then used to drive piezo-electric staks which move to correct the errors. Because this is a close-loop system, non-linearity and hysteresis in the piezo-electric are entirely removed.
The CS100 however can erroneously adjust the geometrical distance of the plates, thus producing wavelength instabilities of the interferometer, due to different effects: i) changes in the dielectric constant of the air, producing a wrong signal from the capacitance micrometers, ii) changes in the spacing between the pads of the micrometers, produced by thermal effects on the capacitor pillars, iii) electronic drifts produced by thermal effects on some components of the CS100.
Moreover, for what concerns the refractive index of the air, this obviously depends on the atmospheric changes of pressure, temperature and humidity.
The environmental effects on both the capacitance micrometers and the optical length of the cavity are easily eliminated by placing the interferometer in a sealed chamber. The residual thermal effects have then been eliminated by accurately thermostatizing both the interferometer and the electronics.
In particular, as the FPI has a thermal sensitivity of
3 mÅ C-1, to obtain a wavelength stability
of
1 ms-1, the interferometer must be thermostatized within
5 10-3
C. To this purpose the
FPI chamber has been enclosed in a copper box,
externally warmed by a sheet resistance, and covered by
heat insulating panels;
two fans mix the air to make the temperature uniform
inside the box, and a thermostatic system allows one to obtain
a thermal stability of
0.1
C at about
37
C.
Then, a second sheet resistance, wrapped around the sealed chamber
and controlled by a second independent thermostatic system,
maintains the interferometer at about 38
C
within
5 10-3
C (Fig. 1 (click here)), for variations of
2
C of the ambient temperature.
Figure 1: Temperature inside the interferometer chamber
vs. time. The temperature has been evaluated by
measuring the resistance of a Pt100, with a resolution of
1 m (
2.5 10-3
C), and the
fluctuations relative to the mean (
38
C)
have been plotted
As regards the CS100, its thermal sensitivity is
0.1 mÅ C-1, and a thermostatization
within
0.1
C, obtained by controlling the
speed of two fans cooling the electronics, is therefore
sufficient to secure the demanded wavelength stability.
Finally, to test the residual drift of the interferometer, the 114Cd
red line at 6438 Å, emitted by an isotopic spectral lamp,
has been used. The stability of this line
(secondary wavelength standard) is 2 10-8
, corresponding to
6 ms-1 in velocity units
(Engelhard & Bayer-Helms 1972).
The following assembly has then been used to perform the test. The interferometer, used in axial mode, was placed between two lenses: the first one collimates a pinhole, where a macrophoto lens forms an image of the lamp discharge tube, the second one forms an image of the pinhole on the cathode of a photomultiplier. A second photomultiplier, used as reference, directly sees, via a fiber optic, the light emitted by the lamp. The photomultipliers are both used in photon counting and are preceded by an interference filter, centered at 6438 Å, sufficiently narrow (FWHM = 50 Å) to isolate the Cd red line.
This assembly was first used to accurately measure the
profile of the Cd line and to locate on it a linear range,
which was found between 60 and 80% of the peak
intensity. The two points, at the center of the linear ranges
on the blue and on the red flank of the line, have been then
located at 13 mÅ from the line peak.
Let us define now as velocity signal the quantity:
with
where ,
and
respectively
are the wavelengths of the line peak and of two points on the
blue and on the red flank of the line;
and
are
the intensities measured on
and
;
and
are the corresponding
intensities measured by the reference photomultiplier.
Fixed
= 26 mÅ,
we have, within the linear range:
K has been then evaluated by a linear fitting of S,
measured at five around
(
= 1 mÅ). The so obtained value for K is:
It follows from Eqs. (13) and (14) that, if the wavelength
stability of the interferometer must be measured within
1 ms-1 (
0.02 mÅ), we must be able
to measure a variation
.
It is easy to show that, assuming
,
the rms errors on S and
respectively are
and
.
If we impose
, we obtain
therefore
107 counts.
To measure this signal in a short time, a photon counting at frequencies of some MHz is obviously necessary. At this rate however the non linearity effects become relevant and the dead time of the chain formed by the photomultiplier, the amplifier-discriminator and the counter must be known. To this purpose, the following second assembly has been used. By means of a beamsplitter cube, the light emitted by a photodiode is seen simultaneously by the two photomultipliers. Between the diode and the cube a shutter allows one to measure the dark counting and a filter wheel, carrying a neutral density (D = 0.1), allows one to reduce the signal of a known amount. For each photomultiplier, the ratio R between the signals with and without density has then been measured at different light levels, up to 6.3 MHz. R has been then reported versus Log I, where I is the signal without density, and the dead time has been obtained by the best fit of the data. The same dead time has been found for the two photomultipliers (38 ns) and, accounting for it, the newly measured R are constant within 0.5%, up to 6.3 MHz.
The first described assembly, used to measure the
profile of the Cd red line, has been then used to measure
the wavelength stability of the interferometer. To this
purpose, the position of the line peak has been located,
and then the velocity signal S has been measured on two
points at 13 mÅ from the line peak and
on two other points at
14 mÅ. By using crossed
polarizers, the signals on the two photomultipliers have
been adjusted to
2 MHz
and an integration time of 5 s has been used, to secure
a sensitivity of 1 ms-1.
The measurement has been repeated each minute for two days
and the results are shown in Fig. 2 (click here), where
the wavelength drift, obtained by Eqs. (13) and (14),
is plotted versus time.
The curves a and b respectively refer to the measurements
performed at
13 and
14 mÅ, and, for the sake
of clarity, they have been arbitrarily shifted along
the y axis. To evaluate the error on the measured drifts,
the curve c has been then obtained as a difference
between a and b.
As it may be seen, c is a horizontal straight line with a superimposed
noise, amounting to
0.85 ms-1.
If we suppose that the error on a and b is similar, we may
conclude that the error on each measured drift is
0.60 ms-1, significatively larger than the
expected error
=
K-1 =
4.9 10-3 mÅ (
0.23 ms-1).
This discrepancy can be explained as due to the electronic
noise of the CS100, which causes the interferometer plates
to make small amplitude random movements about their mean
position. Queensgate claims an electronic noise
equivalent displacement of the FPI plates amounting to
10 pm Hz-1/2 (see Table 1 (click here)).
For a 3 mm separation, this means a noise
ms-1 Hz-1/2,
where t is the integration time. If t = 5 s, the electronic
noise will be therefore
=
0.45 ms-1,
and finally the total expected noise, photometric plus electronic,
will be
0.5 ms-1, very similar to the measured value of
0.6 ms-1.
Figure 2: Wavelength drift of the Fabry-Perot
interferometer vs. time. The curves a and b refer to
independent measurements of the 6438 Å Cd line position,
respectively at 13 and
14 mÅ from
the line peak. For the sake of clarity, the two curves
have been arbitrarily shifted along the y axis.
The curve c, a horizontal straight line with a superimposed
noise of
0.85 ms-1, has been obtained as a difference
between a and b
By carrying on the analysis of Fig. 2 (click here), it may be seen that
the a and b curves
show a very slow, nearly linear trend. From a linear fitting to the
mean of the two curves, a drift of 0.014 mÅ h-1
(0.65 ms-1h-1) is found.
This drift, amounting to 0.67 mÅ (31 ms-1) in 48 h,
cannot be ascribed to the the spectral line,
the stability of which is 6 ms-1,
neither to temperature effects on the interferometer,
as shown by Fig. 1 (click here), but more probably to some electronic
drift of the CS100.
Other shorter measurements of the interferometer wavelength
stability have been performed and different drifts have been
found, but none larger than 5 ms-1 has been
observed over a period of ten hours.
Moreover, the residuals of the linear fitting, shown in Fig. 3 (click here),
evidentiate wavelength instabilities amounting to 0.34 mÅ
(16 ms-1) peak to peak.
The cause of these fluctuations is not evident;
their amplitude, however, very similar to the line instability,
suggests that probably they must be ascribed to the spectral
lamp, rather than to the interferometer itself.
To test if these fluctuations could
be ascribed to the lamp power supply,
the recommended cathod heating current (0.8 A) and the discharge
current (0.3 A) have been changed from 0.6 to 0.9 A and from
0.2 to 0.5 A respectively, but no variation on the velocity signal
within 1 ms-1 has been observed.
It is evident, therefore, that
to have a more accurate knowledge of the instrumental
wavelength drift a more stable source, as, for example, a
frequency stabilized laser, should be used.
Figure 3: Residuals of a linear fitting to the mean of
the a and b curves of Fig. 2 (click here), showing wavelength
instabilities amounting to 0.34 mÅ (16 ms-1) peak to peak
In conclusion, after the inclusion of the interferometer in a sealed chamber and a careful thermostatization, the instrumental profile shows a very high wavelength stability: the maximum expected drift on 10 h, as in the case of a very long solar observation, is about 0.2 mÅ (10 ms-1).
Anyhow, as a measurement of the instrumental
wavelength drift demands a short time to be performed
( 2.5 s for an integration time of 1 s),
such a measurement can be alternated to the observations
and used to compensate the drift, by correcting
the biasing voltage applied to the interferometer.
However, in practice, because the smallest voltage step
allowed by the CS100 corresponds to a wavelength shift of
about 1 mÅ, too large with respect to the observed
drift, this correction will seldomly be useful: when the
instrument does not work correctly.
On the other hand, the measurement of the drift can be also useful a posteriori to verify the instrumental behaviour, and, in case, to correct the data for drifts exceeding the wavelength stability of the Cd lamp.