next previous
Up: The low-energy concentrator

3. Instrument performance

  Great care has been taken to ensure a good calibration of the LECS. Each of the critical components has been separately calibrated and then the entire calibration has been verified at instrument level. The measurements have been performed using SSD's 6 meter X-ray beam, the Berlin synchrotron facility (BESSY), and the PANTER long-beam X-ray facility of the Max Planck Institut für Extraterrestrische Physik near Munich. All measurements reported here were performed in vacuum with the nominal voltages of -20 kV on the cell and 1082 V on the PMT.

Initial functional tests were carried out using SSD's X-ray beam facility. During these tests, the LECS was mounted behind a pin-hole plate and illuminated with Al (1.49 keV) and Fe (6.41 keV) radiation from a source located 5 meters away. The pin-hole plate consists of a 130 tex2html_wrap_inline2354m thick Ni sheet with 1 mm diameter holes in a rectangular grid with a 4 mm pitch. Figure 7 (click here) shows a representative image obtained with Fe Ktex2html_wrap_inline2356 illumination to illustrate the overall performance of the LECS. The two tex2html_wrap_inline2358Fe calibration sources are visible in the upper left and lower right corners. The off-axis pin-hole images are asymmetric and extended towards the outside of the FOV. This is caused by the penetration of tex2html_wrap_inline2360 keV X-rays into the cell gas before absorption. Because of the driftless design (see Sect. 2.2 (click here)), these penetrating X-rays will give rise to lower PHA and BL values than those absorbed directly beneath the entrance window. In addition, the variation in viewing angle and footprint of the scintillation light will cause the measured position to be shifted. The dependence of mean penetration depth on energy under the LECS nominal operating conditions is shown in Fig. 8 (click here).

  table365
Table 2: PANTER calibration lines

  figure380
Figure 7: Multi pin-hole array 6.41 keV image. The calibration sources are the bright spots in the upper left and lower right corners. The circles indicate the regions used for gain mapping. The circular FOV of the LECS is approximately bounded by the outermost pin-holes

  figure385
Figure 8: Mean absorption depth in Xe as a function of X-ray energy using the absorption coefficients of Henke et al. (1993). The positions of the Xe absorption edges and the energies of the lines used in the LECS PANTER calibration are indicated

The effect of penetration on the LECS event data is further illustrated in Fig. 9 (click here). This shows BL versus energy for incident X-ray energies just above and just below the Xe tex2html_wrap_inline2388 edge, where there is a large discontinuity in mean X-ray absorption depth (see Fig. 8 (click here)). At an energy of 4.77 keV (i.e. just below the Xe tex2html_wrap_inline2390 edge) the mean penetration depth is large and the extended tail of events in the upper panel of Fig. 9 (click here) towards lower energy and BL values (1) is caused by penetration. The events labeled as (2) are due to multiple events occurring within the EU sampling window i.e. event pile-up. This feature is prominent due to the high count rates used for these calibration measurements, but will be negligible for most astronomical sources. In contrast, the lower panel of Fig. 9 (click here) is an exposure at 4.79 keV, i.e. just above the edge. As expected, the effects of penetration are greatly reduced with only a small tail towards lower energies and BL values evident (1). In addition the fraction of escape photons (3) has increased due to the incident photons having sufficient energy to liberate electrons from the Xe tex2html_wrap_inline2392 shell. A further discussion of these effects can be found in Simons & de Korte (1989) and Bavdaz et al. (1995a).

  figure399
Figure 9: BL versus PHA channel for 4.77 keV (upper panel) and 4.79 keV photons (lower panel), i.e. just below and above the Xe tex2html_wrap_inline2394 edge. See the text for a description of the labels

3.1. Gain temperature dependence

  A primary concern is to understand the ``gain'' of the instrument. This is the relation between the energy of an event and the mean PHA channel at which it is detected. The instrument functional tests were performed over a range of temperatures allowing the temperature dependence of the instrument performance to be investigated. The DU thermistor is used as temperature sensor, since this will be the in-orbit reference.

Each instrument functional test was divided into intervals of constant temperature and for each of these segments the average PHA channel of each calibration source was determined (see Fig. 10 (click here)). The results from the two calibration sources only differ by a scaling factor which probably arises from a small misalignment of the PMT and gas cell fundamental axes. Both calibration sources show a linear dependence of gain with temperature in the range studied. The best-fit coefficient is -1.25% tex2html_wrap_inline2406 at a PHA channel of 410. The scatter in the figure can be attributed to the 0.4 tex2html_wrap_inline2408 resolution of the temperature measurement in the region around tex2html_wrap_inline2410. The same procedure was performed using Fe tex2html_wrap_inline2412 and Al tex2html_wrap_inline2414 data from within the FOV to confirm that the temperature dependence is independent of both position and energy. The temperature dependence of the gain can be corrected for by defining a nominal calibration source PHA channel of 410 and normalizing all energy signals to this. If the original energy channel = PHA, then the corrected channel is given by:
 equation411
where caltex2html_wrap_inline2416 and caltex2html_wrap_inline2418 are the mean PHA channels of the two calibration sources. Note that the nominal calibration source channel of 410 is that expected at the predicted in-orbit LECS temperature of tex2html_wrap_inline2420.

  figure419
Figure 10: PHA channel versus temperature for the two tex2html_wrap_inline2422Fe calibration sources. The dotted lines show the best-fit linear relations

3.2. Gain position dependence

  Spatial gain maps were obtained by computing the gain at each of the pin-hole locations indicated by circles in Fig. 7 (click here). The gain of each image pixel was then estimated by spline interpolation. Temperature effects were removed by dividing the data into intervals of constant temperature and using the average calibration source PHA to normalize the amplitude of each individual X-ray event according to Eq. (5 (click here)). The peak PI channel was determined by fitting a model consisting of a Gaussian profile plus a constant background to the data. The interpolated gain maps are shown in Fig. 11 (click here) for Fe tex2html_wrap_inline2426 and for Al tex2html_wrap_inline2428 X-rays. The region of maximum gain is offset slightly from the center of the FOV, probably because of the small misalignment between the PMT and gas cell fundamental. At the edge of the FOV the gain is tex2html_wrap_inline2430% of the maximum. The two gain maps only differ by a scaling factor of 0.238, consistent with the Al/Fe X-ray energy ratio of 0.232 and the energy discontinuities at the Xe L edges.

Additional multi pin-hole measurements were taken at the PANTER facility, where other X-ray energies are available. During this campaign, the pin-hole array was stepped in the XY plane, parallel to the focal plane, to increase the spatial sampling. These measurements confirm the smooth surfaces and extend the calibration to lower and higher energies. Again, the spatial gain maps are well behaved, differing only by a scaling factor consistent with the X-ray energy ratios and detector gas edge discontinuities.

  figure429
Figure 11: Gain maps at 1.49 and 6.41 keV. The crosses indicates the center of the FOV. The numbers indicate the gain relative to the average of the calibration sources

3.3. Penetration correction

  Due to the driftless design of the gas cell, X-rays that penetrate into the cell prior to absorption give rise to events with lower PI and BL values than events of the same energy which are detected just beneath the cell entrance window (see Figs. 9 (click here) and 12 (click here)). In order to correct the observed PI spectra for this effect, data obtained at the PANTER facility have been used to determine the mean PI channel as a function of BL for incident energies tex2html_wrap_inline2440. (Below this energy there is no worthwhile improvement in spectral resolution when this correction is applied). For each line a reference PI channel, tex2html_wrap_inline2442, was determined. This corresponds approximately to the mean channel of events absorbed directly beneath the cell entrance window. In order to determine tex2html_wrap_inline2444 for each line, the function described in Eq. (8 (click here)) together with a constant background were fit to the BL spectra and events with BL values greater than the peak of the distribution selected. A PI spectrum of these selected events was then accumulated and the same function fit in order to account for the remaining small penetration tail. The value of tex2html_wrap_inline2446 is then given by the best-fit value of m. Thus, for a known energy, tex2html_wrap_inline2450, the observed PI spectrum can be corrected for penetration using:
 equation445
Here PIC is the corrected channel and tex2html_wrap_inline2452 is the mean PI channel for the observed BL value and incident energy. From an examination of the tex2html_wrap_inline2454 distributions it is apparent that the BL resolution of the instrument is insufficient to fully correct for penetration effects in the energy range of interest. This explains the remaining small exponential tail visible in the corrected spectrum in the lower panel of Fig. 12 (click here).

During normal astronomical observations there is a further complication that the actual energy of a detected event, tex2html_wrap_inline2456, is not known a priori and only the probability that an event has a certain incident energy can be estimated. In the LECS data analysis software a value of tex2html_wrap_inline2458 is assigned to each event taking into account this probability. The penetration correction can then be applied using Eq. (6 (click here)). This lack of a priori knowledge of the incident energy means that e.g. the measured FWHM energy resolution of 9.0% at Fe tex2html_wrap_inline2460 (6.41 keV) can only be corrected to 8.6% using this technique, rather than 8.1% if the energy of the incident X-rays is known.

3.4. Energy to channel conversion

  The overall energy gain of the LECS is expected to be linear, with discontinuities at the absorption edges of the detector gas caused by discrete changes in the final ionization state, and hence the amount of energy retained by the Xe atoms across the absorption edges. In order to measure the overall gain, results from the PANTER measurement campaign were utilized since these covered almost the entire energy range of the instrument. The lines used and their energies (Bearden 1967) are given in Table 2 (click here). For each of these lines, the PHA data were first corrected for the overall temperature and position dependence of the gain using Eq. (5 (click here)) and the mean gain map (see Fig. 11 (click here)).

  figure470
Figure 12: The response of the LECS to monochromatic radiation at energies of 0.28 keV (upper panel) and 8.06 keV (center panel). The statistical uncertainties are comparable to the symbol size. The ``knee'' visible in the tex2html_wrap_inline2462 spectra at channel tex2html_wrap_inline2464 is from Cu tex2html_wrap_inline2466 radiation. The lower panel shows the effect of performing the penetration correction described in Sect. 3.3 (click here) on the Cutex2html_wrap_inline2468 spectrum assuming an incident energy of 8.06 keV

The resulting spectra, tex2html_wrap_inline2470, can either be represented by a Gaussian or, at energies tex2html_wrap_inline2472 keV, a Gaussian plus an exponential low-energy tail. The upper two panels in Fig. 12 (click here) show observed line profiles at 0.28 keV, where the mean absorption depth is small, and 8.06 keV, where it is large (see Fig. 8 (click here)). The 0.28 keV line profile is well fitted by a Gaussian function, demonstrating the good low-energy performance of the instrument since the loss of significant numbers of electrons would result in a tail towards low energies, which is not evident. The 8.06 keV line shows an exponential tail extending towards lower energies. This results from the penetration of X-rays into the gas cell prior to absorption. The observed line profiles were modeled using a constant background together with the following functions:

where tex2html_wrap_inline2476 is the break channel between the two components, tex2html_wrap_inline2478 is the e-folding index, m is the peak PI channel of the Gaussian, tex2html_wrap_inline2482 is the width of the Gaussian distribution in standard deviations and tex2html_wrap_inline2484 and tex2html_wrap_inline2486 are the normalizations of the exponential and Gaussian components, respectively. Since the two components connect smoothly at tex2html_wrap_inline2488, their first derivatives are equal at this point. (This implies that tex2html_wrap_inline2490). Some idea of the quality of fits can be obtained from the upper two panels in Fig. 12 (click here) where the fitted models are shown as continuous lines. The penetration correction given in Eq. (6 (click here)) was then applied to the data from all lines with energies > 2.9 keV (below this energy the PI and PIC channels are equivalent). Equations (7 (click here)) or (8 (click here)) were then fit to the PIC spectra, as appropriate, to derive the best-fit values of m which are used for the PIC to energy conversion. A linear relation between PIC value and energy was assumed, allowing for a discontinuity at the Xe tex2html_wrap_inline2496 edge at 4.782 keV. The result of this fitting gives:


equation512

The discontinuity at the Xe L edge is 31 eV. This value is smaller then the 129 eV measured by Bavdaz et al. (1995a) using a prototype detector at BESSY. This difference probably arises because these authors selected only a small fraction of the available data to minimize penetration effects. Such a technique cannot be usefully applied to astronomical measurements. No evidence for a gain discontinuity at the Xe M edge was found in the PANTER data.

3.5. Energy resolution

  The energy resolution of the LECS was determined at each of the energies listed in Table 2 (click here) using either Eqs. (7 (click here)) or (8 (click here)) and corrected for penetration effects using Eq. (6 (click here)), as appropriate. No a priori knowledge of the line energy was assumed, since this will be the case during astronomical observations. The full width at half-maximum (FWHM) of the PIC energy resolution, tex2html_wrap_inline2504, is related to the width of the Gaussian distribution by tex2html_wrap_inline2506. The LECS energy resolution is given by:
 equation530
and is shown plotted versus energy in Fig. 13 (click here). The deviation from the expected tex2html_wrap_inline2508 relation has been previously noted (Simons & de Korte 1989) and explained as being due to variations in the footprint of the electron cloud viewed by the PMT.

  figure541
Figure 13: The LECS penetration corrected energy resolution versus energy. The best-fit relation given in Eq. (10 (click here)) is shown as a solid line. The uncertainties are smaller than the symbol size

3.6. Image linearization

  The coordinate transformation required to translate the RAWX, RAWY pixel coordinates into absolute position on the focal plane were obtained using multi pin-hole mask (MP) exposures taken at PANTER. Due to the finite source distance of 130 m the actual distance between consecutive pinholes projected on the focal plane is magnified by tex2html_wrap_inline2512%. This effect is taken into consideration in the analysis. The data used for this analysis were obtained using lines of C (0.28 keV), Al (1.49 keV) and Fe (6.41 keV). For each exposure, the relation between pixel coordinates and pin-hole position, as shown in Fig. 14 (click here), was established.

Based on this data, the coordinate transformation required to assign to each RAWX, RAWY pixel its true distance along the X and Y axes from the physical center of the FOV have been determined. A cubic polynomial function has been fit to the data:
 eqnarray549
Here, tex2html_wrap_inline2518 is the distance in mm from the physical center of the FOV (which is located at pixel coordinates RAWX, RAWY = 131.44, 124.12) along the X and Y axes, tex2html_wrap_inline2524, tex2html_wrap_inline2526 and tex2html_wrap_inline2528 are the fitted coefficients (one set for each axis). The rms residuals between the data and the fit along both axes are tex2html_wrap_inline2530. This corresponds to about half a pixel and is of the order of the systematics inherent to the experimental set up. From a comparison of the pin-hole images taken at different energies, any energy dependence corresponds to a positional difference of tex2html_wrap_inline2532, except at the edge of the FOV where these may be as large as tex2html_wrap_inline2534. The positions in mm are converted to linearized pixels by the LECS data analysis software assuming a pixel size of 8''.

  table569
Table 3: Coefficients required to map pixel coordinates into absolute position on the focal plane

  figure609
Figure 14: Correspondence between tex2html_wrap_inline2602, tex2html_wrap_inline2604 pixels (left panel) and physical space (right panel). Each data point represents the position of one pinhole

3.7. Encircled energy function

  The overall encircled energy function (EEF) of the LECS is a convolution of those of the MU and detector modulated by the obscuration of the entrance window strongback and support grid. Its shape can therefore not be easily parameterized in a similar manner to that of the MECS (Boella et al. 1996a). Measured, azimuthally averaged, on-axis EEFs are shown in Fig. 15 (click here). At low energies the EEF is dominated by the contribution of the detector, which is approximately tex2html_wrap_inline2608. This results in the narrowing of the EEF core visible in Fig. 15 (click here) with increasing energy. Above tex2html_wrap_inline2610 keV, the broad scattering wings of the mirror response become important (see Fig. 19 (click here)). The FWHM of the LECS EEF is given by:
equation621
At 6 keV this relation is valid in the central 8' of the FOV. The range of validity increases with decreasing energy, such that at 0.28 keV it is valid over the entire FOV.

  figure628
Figure 15: Measured, azimuthally averaged, on-axis LECS encircled energy functions at various energies. Each plot has been offset from the next by 0.1 for clarity. The dotted lines show the asymptotic values

3.8. Effective area

  The effective area of the LECS is primarily limited at low-energies by the transmission of the entrance window and at high-energies by the loss in reflectivity of the MU. The predicted on-axis effective area of the LECS is shown in Fig. 16 (click here). This figure includes the theoretical mirror efficiency (Sacco 1995, private communication), detector absorption efficiency and plasma grid transmission and the measured entrance and plasma protection window transmissions. A constant geometric obscuration due to the window support structure of 17% is assumed. The large change in effective area at 0.28 keV is due to the C K-edge. The effect of the K-edges of N, O, and Al, which are all constituents of the windows, are also visible. The structure between 2-4 keV is caused by the M-edges of the Au mirror material. At the edge of the FOV (i.e. an offset angle of 18'), the effective area of the LECS is 0.4 times that on-axis.

  figure637
Figure 16: Predicted LECS on-axis effective area

3.8.1. Mirror effective area

  In the PANTER configuration the effective mirror collecting area is reduced to tex2html_wrap_inline2626, including obscuration by the mirror spider (Conti et al. 1994). The MU effective area was measured at each of the energies listed in Table 2 (click here) (except that the measurements for the two Ag and Ti lines were combined, and an additional P (2.01 keV) measurement was included). The MU effective area was derived by comparing the count rate with the MU in place to that during a flat-field exposure (i.e. with the MU removed). The derived values were corrected for the effects of strongback and fine grid obscuration and deadtime and are shown plotted in Fig. 17 (click here). This figure also shows the theoretical MU response derived from ray-tracing simulations of a source at the same distance as that at PANTER. The agreement between measurement and theory is good except at low energies (the B and C measurements at 0.18 and 0.28 keV) where the measured areas are a factor tex2html_wrap_inline2628 too high. Such an effect may be related to the location of the LECS in the PANTER chamber and the complicated behavior of the MU to off-axis X-rays. In particular, X-rays from within a tex2html_wrap_inline2630 offset angle which are reflected by either one of the first or second mirror cones only, can reach the focal plane. In addition, for particular offset angles, X-rays can reach the focal plane without reflection (Conti et al. 1994). Thus, it is possible that low-energy X-rays scattered off the chamber walls during the mirror exposures enter the detector, resulting in the anomalous area measurements. An additional effect may be due to molecular contamination of the mirror surfaces (e.g., Elsner et al. 1994). However, there is no totally convincing explanation for this phenomenon and we await the results of in-flight measurements to verify the low-energy LECS effective area.

  figure649
Figure 17: Measured and theoretical (solid line) on-axis MU effective area versus energy in the PANTER configuration

3.8.2. Strongback and grid obscuration

  The construction of the strongback and grid are shown in Figs. 4 (click here) and 5 (click here), respectively. The strongback and grid together obscure an average of 17% of normally incident X-rays. However, this obscuration depends on both position in the FOV and X-ray energy in a complex way. The energy dependence arises because:

  figure664
Figure 18: A scan across the central part of the LECS window centered on 0.28 keV showing the effects of strongback and fine grid obscuration. The distance scale indicates the approximate angular distance from the center of the FOV

The effect of strongback and fine grid obscuration is illustrated in Fig. 18 (click here) which shows the LECS counting rate at 0.28 keV during a 3.9 mm long scan of the central region of the detector in a direction parallel to the Y axis, using a constant intensity X-ray beam. The two large (tex2html_wrap_inline2636%) decreases in count rate delineate the 2.2 mm central strongback square and are caused by the strongback ribs. A smaller amplitude (tex2html_wrap_inline2638%) modulation due to the fine grid is also visible. The overall trend of increasing count rate with distance along the Y axis is probably due to a small misalignment between the mirror and detector units.

These obscuration effects are treated by means of a Monte-Carlo simulation in the LECS data analysis system. The first stage of this process is a ray-tracing model of the mirror system. This is based on the commercial ray tracing software package, IRT, from Parsec Technology Inc. which has been ``tuned'' to give good agreement with the mirror EEF's obtained at the PANTER facility using the flight MU and the ROSAT Position Sensitive Proportional Counter (PSPC) as a focal plane detector. The spatial resolution of the PSPC is much better than that of the LECS, and is sufficient to allow the intrinsic shape of the MU EEF to be reliably determined. Figure 19 (click here) shows the good agreement between observed MU EEFs and those predicted by the ray-tracing model.

A geometric model of the support structure and the fine grid is then used to determine the fate of an X-ray with given energy and arrival direction. X-rays which exit the MU can reach the entrance window undisturbed, or be absorbed in the fine grid, or interact with the strongback. In the latter case, the photon can be either absorbed or reflected, with a probability that depends on the energy and the angle of incidence with the strongback. Reflected photons may be absorbed by the fine grid. Using the processes described above, it is possible to generate the EEF of the concentrators for any given energy and position in the FOV and to propagate this through the support structure. The effect of this obscuration on the analysis of both extended and variable sources will be evaluated during the Science Verification Phase of the mission.

  figure674
Figure 19: Azimuthally averaged mirror unit EEFs at two energies obtained by ray-tracing (squares) and measured using the ROSAT PSPC at two energies (solid lines). The plots have been offset by 0.5 units for clarity. The dotted lines show the asymptotic values

Entrance window X-ray transmission

  The primary goal of the LECS is to perform low-energy spectroscopy. Since the low-energy response of the detector is primarily defined by the transmission of the entrance window, a well calibrated window transmission is vital. With this in mind, great care was taken to determine the window transmission paying particular attention to variations near the edges of the constituent materials and to the position dependence of the overall transmission.

Measurements using the available lines at PANTER (Table 2 (click here)) are inadequate for this purpose, and instead a thorough investigative campaign was performed at BESSY. The PTB SX-700 plane grating monochromator beamline was used which provided energy resolution, tex2html_wrap_inline2646, of between 5000 at 0.06 keV to 400 at 1.8 keV, scaling as tex2html_wrap_inline2648. Since the entrance window is highly transparent at energies tex2html_wrap_inline2650, measurements were only made in the energy range 0.1 to 1.8 keV. Transmissions at 500 energy values were measured at five positions in the central window area. Energy steps as small as 0.2 eV were used around the absorption edges of the constituents. These measurements were complemented by a series of measurements in the center of each of the 8 by 8 fine grid positions in the central 3 by 3 central strongback squares (see Fig. 4 (click here)). Between 18 and 30 energies were used for these scans. The 99% flux width of the beam was tex2html_wrap_inline2652, smaller than the fine grid mesh size. This allowed the transmission of the window to be measured separately from the obscuration caused by the support structure. The absolute errors on the transmission are tex2html_wrap_inline2654% and the variations in transmission with position < 3%. A detailed discussion of these measurements is to be found in Bavdaz et al. (1994, 1995b).

The X-ray cross sections used by Bavdaz et al. (1994) were derived from tables in Veigel et al. (1971). Another, more recent, set of coefficients has been published by Henke et al. (1993). They differ from the older ones mainly around the absorption edges and are used for the LECS calibration. The usual approach to modeling X-ray transmission, T, through an absorbing layer is to use:
equation691
where tex2html_wrap_inline2660 is the density, x the thickness and tex2html_wrap_inline2664 the mass absorption coefficient, which for compounds is given by:
 equation693
where tex2html_wrap_inline2666 is the mass absorption coefficient of element i and tex2html_wrap_inline2670 its fractional weight. By defining tex2html_wrap_inline2672 as the area density of element i, we can write:
equation702
Using the measurements at different energies, gives the area densities required to describe the X-ray transmission properties of the window. Unfortunately the chemical state and thickness of the absorber change the edge absorption characteristics e.g., Bearden & Burr (1967) and Owens et al. (1996) and Eq. (14 (click here)) is only a good approximation away from the edges. As an example, Fig. 20 (click here) shows the actual transmission around the O K edge compared to the Veigel et al. (1971) (solid line) and Henke et al. (1993) mass absorption coefficients (dashed line).

In order to precisely model the LECS window transmission, the following empirical function has been used:
equation719
where the correction factor, tex2html_wrap_inline2676, is defined as the ratio of the measured data to that predicted using the Henke et al. (1993) coefficients outside the edge regions:
equation727
The correction factor is derived from the average of the five high resolution energy scans and is plotted in Fig. 21 (click here). As expected, it has a value close to one (i.e. no correction) away from the edges of constituent materials and above 1.8 keV.

  figure735
Figure 20: Entrance window transmission around the O K edge. Predictions using theVeigel et al. (1971) and theHenke et al. (1993) mass absorption coefficients are shown as solid and dashed lines, respectively. The measured data are shown as crosses

  figure741
Figure 21: The correction factor used to model the transmission of the entrance window around the absorption edges

The area densities for each of the window constituents were fit at each position by minimizing the parameter, Q:
eqnarray746
where the index ``j'' denotes the energy dependence and ``k'' the window constituents of H, C, N, O, and Al. Energy regions close to the absorption edges were excluded. Table 4 (click here) summarizes the theoretical composition, based on the design shown in Fig. 3 (click here). Figure 22 (click here) shows the measured transmission of the entrance window and compares it with a tex2html_wrap_inline2684 thick Be window as used on the ASCA GSPC (Tanaka et al. 1994). The derived area densities are given in Table {5 (click here). The fitted curve matches the data well, but is not superposed for clarity. The complex structure near the edges is clearly visible.

  figure764
Figure 22: The measured transmission of the LECS entrance window at one position. Absorption edges of the constituent materials are indicated. The theoretical transmission of a tex2html_wrap_inline2686 Be window is also shown for comparison

  table771
Table 4: Nominal entrance window composition

  table789
Table 5: Measured entrance window composition

Plasma protection window X-ray transmission

  The transmission of the plasma protection window was also measured at BESSY. A high resolution energy scan (500 energy measurements between 0.1 and 1.8 keV) at the center of the window and a high position resolution scan at 3 energies (0.27, 0.90 and 1.55 keV) were performed. The scan revealed < 1% rms spatial variations in transmission. Since this window is located 5 cm above the focal plane, any small transmission variations will be smoothed out and we therefore assume a uniform absorption with position. The absorption of the window has been modeled in a similar manner to that of the entrance window, except that the effect of the support grid is included. The resulting function is plotted in Fig. 23 (click here).

  figure818
Figure 23: The measured transmission of the plasma protection window and strongback. Absorption edges of the constituent materials are indicated

3.8.5. Detector absorption efficiency

  Nearly all incident X-rays with energies in the range 0.1-10 keV that pass through the entrance window are absorbed in the 5 cm deep LECS gas cell. The absorption efficiency decreases slightly at the upper end of the energy range, being 0.996 at 10 keV. Following photo-electric absorption an excited ion may relax by the emission of an Auger electron or a fluorescent photon. The probability that a fluorescent photon will escape from the detection volume is geometry dependent, and has been measured for the LECS using data obtained at BESSY and PANTER. The values given in Table 6 (click here) have been adopted for use in the LECS calibration and show that 1.68% of incident X-rays with energies just above the Xe tex2html_wrap_inline2706 edge produce fluorescent X-rays which are lost from the instrument, resulting in escape peaks. This figure is less than the Xe L shell fluorescent yield of tex2html_wrap_inline2710% due to the probability that an emitted fluorescent photon is absorbed within the detector. Table 6 (click here) gives the escape fractions for energies between the Xe tex2html_wrap_inline2712 and Xe tex2html_wrap_inline2714 edges. Above the Xe tex2html_wrap_inline2716 edge the escape fraction is assumed to decrease linearly with energy, falling to 1.12% at 8.06 keV.

  table833
Table 6: LECS escape line data

3.9. Background counting rate

  The VETO and BL signals can be used to minimize the residual background counting rate seen in the LECS. The definition of these signals is given in Sects. 2.2 (click here) and 2.3 (click here). Briefly, the VETO signal is a measure of the fraction of the total light collected by the central anode of the PMT (see Fig. 6 (click here)) and, in general, an on-axis event will have a higher value than an off-axis or extended event. The BL signal is a measure of the duration of scintillation of an event and will, on average, be shorter for events that penetrate into the detector prior to absorption.

  figure867
Figure 24: BL plotted against VETO at two energies

Figure 24 (click here) shows BL plotted against VETO for on-axis B (0.18 keV) and Cutex2html_wrap_inline2742 (8.06 keV) events. Events falling outside the FOV are excluded. The events with VETO values of tex2html_wrap_inline2744 result from multiple calibration source events being detected during a single EU sampling window. The effects of penetration are clearly seen in the Cutex2html_wrap_inline2746 image, where a tail towards lower BL and higher VETO values is evident. This change in VETO value is caused by the increased PMT viewing angle of penetrating events which are absorbed closer to the central anode (see Fig. 6 (click here) and Eq. (4 (click here))). At 0.18 keV, because of low signal strength, the BL signal distribution is significantly broadened by the EU circuitry. In addition, at low energies there is an unexpectantly large number of events with high BL values. The magnitude of this effect appears to be inversely correlated with mean absorption depth. It is possible that these events are absorbed very close to the entrance window where the electric field is non-uniform due to the undulating window surface. Figure 25 (click here) shows an optical image of part of an entrance window foil and illustrates the irregular nature of the foil in each of the 250 by tex2html_wrap_inline2748 squares which correspond to the gridlets of the window support grid (see Fig. 5 (click here)).

  figure880
Figure 25: An optical image of part of an entrance window foil showing the irregular nature of the foil in each of 250 by tex2html_wrap_inline2750 gridlets. The lightly colored thin lines show the location of part of the window support grid. Part of a strongback rib is located over the thicker line

  figure885
Figure 26: The allowed BL range for 90% X-ray acceptance efficiency shown as a hatched region

Figure 26 (click here) shows the allowed range of BL plotted against PI energy channel for a 90% X-ray acceptance as a hatched area. The broadening of the BL distribution at low-energies discussed above is visible, as is the effect of change in mean absorption depth around the Xe L edge at a PI channel of tex2html_wrap_inline2752. This figure has been obtained using PANTER data for the lines shown in Table 2 (click here) (except that the results for the Ag tex2html_wrap_inline2754 and tex2html_wrap_inline2756 and Ti tex2html_wrap_inline2758 and tex2html_wrap_inline2760 lines have been combined) and by making a series of ``box cuts'' centered on the calibration energies.

The LECS background counting rate within the FOV has been estimated using a tex2html_wrap_inline2762 exposure taken in SSD's X-ray beam. A VETO acceptance interval of 30 to 65 combined with BL cuts of the type described above were applied to the data to give count rates of tex2html_wrap_inline2764 for 90% X-ray acceptance and tex2html_wrap_inline2766 for 99% X-ray acceptance. (Note that the tex2html_wrap_inline2768 in the count rate units refers to the window area). Figure 27 (click here) shows the effect of these selections. The increase in counts observed in the incident spectrum around PI channel 600 is primarily due to event pile up from the calibration sources. As Fig. 27 (click here) illustrates, these events are efficiently removed by the VETO selection. At the lowest PI channels there is an increase in detected event rate which is probably due to electronic noise. The EU analog thresholds were modified following the PANTER calibration and this effect is not expected to occur in space.

  figure903
Figure 27: The effect of VETO and BL selection on the laboratory background spectrum within the FOV. The observed spectrum is shown as a dashed histogram and the spectrum remaining following selection as a solid histogram. 1tex2html_wrap_inline2770 uncertainties are indicated. The count rate scale is given by the right ordinate. The efficacy of each of the selection techniques is given by the left ordinate. See the text for a description of the spectra


next previous
Up: The low-energy concentrator

Copyright by the European Southern Observatory (ESO)
web@ed-phys.fr