One method has been suggested to overcome these complications in the context of liquid xenon time projection chambers (Aprile et al. 1993). Here, this method is formalized as Compton Kinematic Discrimination (CKD) and examined in more detail. This technique allows the order of the photon interactions to be determined with high probability, as well as providing the basis of a powerful tool for background suppression in GCTs.
CKD takes advantage of redundant measurement information in an event
to determine the most likely interaction sequence. A photon of initial energy E
(using the notation in Sect. 2) interacts in the instrument at N sites,
depositing an energy of Ei at each location
.
It is assumed that the
interactions
are Compton scatters, and interaction N is the final
photoabsorption. Given the correct ordering of the interactions, there are two
independent ways of measuring N-2 of the
scattering angles,
.
Geometrical measurement of
.
From simple vector analysis, given
the correct ordering of the interaction sites one can derive the scatter angles
![]() |
(9) |
![]() |
(10) |
Compton kinematics measurement of
.
Given the correct ordering,
the measured values of Wi can be derived, which were defined earlier as the
energy of the photon after each scattering i, in units of
.
The Compton
scatter formula (Eq. 2) gives:
![]() |
(11) |
![]() |
(12) |
As a first test, trial orderings that produce values of
are ruled out,
since
for any scattering angle
.
This condition will eliminate
many orderings which cannot physically be due to multiple Compton scatters
followed by photoabsorption. Next a least-squares statistic measuring the
agreement between the redundant scatter angle measurements is defined:
![]() |
(13) |
This consistency statistic also provides a powerful tool for rejecting
background events. If the event is truly a multiple Compton scatter event
followed by a photoabsorption then
.
By setting a maximum
acceptable level for
,
events that do not fit this scenario can be
rejected. Such events include partially-deposited photons which scatter out
of the instrument (Compton continuum), photon interactions with spatially
unresolved interaction sites, events with interactions below the detector
threshold, pair-production events, and similarly
decays. These events
frequently have
,
allowing a strong rejection statistic that is not very
sensitive on the level set on
.
Here,
has been treated
as a normal least-squares statistic with N-2 degrees of freedom, and events are
rejected which have probabilities of
.
Variations in the level between
and
do not strongly
affect CKD rejection capabilities.
For example, varying this level from
to
shifted the CKD efficiency
curves in Fig. 4 by 1-2%.
The fraction of 3+ site photopeak events which have the first and second
interaction sites spatially resolved - and hence could be imaged to the proper
direction - is shown in Fig. 4 as a function of energy, for the instrument
model discussed in Appendix A.
Roughly
of all events from 0.2-20 MeV have their first and second sites
spatially resolved from each other.
(Some of these events do not have their second, third, etc.,
interactions spatially resolved from each other, and will be
rejected by the limits on
.)
This figure also shows the fraction of the photopeak events which
CKD properly orders (correctly reconstructed),
as well as the fractions improperly ordered (hence
incorrectly imaged to off-source background), and the fraction completely rejected. For
energies below
10 MeV, CKD allows proper reconstruction (hence imaging)
of
of the photopeak events, while rejecting
.
The remaining
are incorrectly imaged into the off-source background. For
comparison, if the order of the interaction sites were randomly chosen
would be correctly imaged, while the remaining
would be incorrectly
imaged into the background.
Copyright The European Southern Observatory (ESO)