The polarimetric observations were carried out during several runs at
the European Southern Observatory (ESO La Silla, Chile) from 1996 to
1999, using the 3.6 m telescope equipped with the EFOSC camera and
spectrograph. In 1996, the instrument was EFOSC1 and the detector a
512512 TeK CCD (ESO#26) with a pixel size of 27
m
corresponding to 0
605 on the sky (Melnick et al. 1989). Later,
EFOSC1 was replaced by EFOSC2, and the detector was a 2048
2048
Loral/Lesser CCD (ESO#40) with a pixel size of 15
m corresponding
to 0
157 on the sky (Patat 1999). The latter CCD was used
in a 2
2 binning mode, except in October 98.
With EFOSC, polarimetry is performed by inserting in the parallel beam a
Wollaston prism which splits the incoming light rays into two
orthogonally polarized beams. Each object in the field has therefore two
images on the CCD detector, separated by 10
or 20
(depending on the Wollaston), and orthogonally polarized. To avoid image
overlapping, one puts at the telescope focal plane a special mask made
of alternating transparent and opaque parallel strips whose width
corresponds to the splitting. The object is positioned at the centre of
a transparent strip which is imaged on a region of the CCD chosen as
clean as possible. The final CCD image then consists of alternate
orthogonally polarized strips of the sky, two of them containing the
polarized images of the object itself (di Serego Alighieri 1989, 1998;
Lamy & Hutsemékers 1999). Note that the polarization measurements do
not depend on variable transparency or seeing since the two orthogonally
polarized images of the object are simultaneously recorded. The
20
Wollaston was used during the September 96, April 98, and
September 99 observing runs, while the 10
Wollaston was used in
October 98 and April 99.
In order to derive linear polarization measurements, i.e. the two
normalized Stokes parameters q and u, frames must be obtained with
at least two different orientations of the Wollaston. With EFOSC1, this
was done by rotating the whole instrument by 45
(usually at the
adapter angles 270
and 225
)
such that, for each object,
two frames were secured (Melnick et al. 1989). The excellent
transmission of the Wollaston makes these two orientations sufficient
(di Serego Alighieri 1989). With EFOSC2, a half-wave plate (HWP) is
inserted in the optical path and four frames with the HWP at position
angles 0
,
22.5
,
45
,
and 67.5
were obtained,
without the need of rotating the whole instrument (Schwarz & Guisard
1995).
Most observations were done with the Bessel V filter. A few additional
ones were obtained with the Gunn i filter. Typical exposure times are
around 5 minutes per frame. Seeing was around 1
2 except in
September 99 (
). In addition, polarimetric calibration
stars were observed in the same filters in order to unambiguously fix
the zero-point of the polarization position angle and to check the whole
observing and reduction process. The observed standard stars are given
in Table 1.
Copyright The European Southern Observatory (ESO)