Up: First results from THEMIS mode
Subsections
Several tests have been performed to assess the noise in the data. We have
measured three different types of
noise. None of them includes systematic errors. We anticipate that
with an improved setup those
errors will disappear, or at least will be greatly diminished in the future.
They have been discussed in previous
sections: differences in the equivalent widths for
each path, etc. The noise discussed here is that which remains if those
systematic errors were not present.
We have measured the photon noise per pixel using the continuum region around
the Fe I line at 5576 Å (data obtained on August 23rd, see Table 1). One
hundred
images of the disk centre for the Stokes V parameter were obtained at around
7:50 UT. The best way to measure photon noise is to compare two images
identical except for the noise. As an approximation to this ideal we compared
each image with the next one. But solar surface details were visible in the
images,
therefore we created two mean images by adding the odd and even numbered
images respectively. The final images are free from solar features and
comparison
is done between the nearest images possible. A mean
profile of the selected continuum region was taken for each mean image by
adding the profiles along the slit. At this point, each one of the mean
profiles contains
(i=1,2): a common signal Iand an added noise
.
The ratio of the two profiles gave
us (neglecting second order terms):
The mean value of this ratio is therefore 1 if the noise is well distributed
around zero. The standard deviation of the ratio corrected for the number of
images and the number of profiles along the slit (used to generate the mean
profile) gives the noise per pixel. The resulting measured value is
.
This value is consistent with a constant of 250 e-/ADU
(Analog--Digital Unit), as defined by the CCD's chipset constructor.
Next, we searched for the error in the continuum region in the reduced images
of Stokes I, Q, U and V. The same 100 images of the disk center at 5576 Å
were used for this purpose, but after reducing them as explained in previous
sections and obtaining the Stokes parameters by simple addition and
substraction. Each set of profiles from the four resulting images for the
Stokes
parameters was interpreted as a sole statistical variable. No consideration
was made about how they had been obtained. The continuum in Q, U, V should
show a null
signal plus an error. In practice we observed a mean signal of 2 10-4indicating a problem with the normalization of the continuum signal from the
two paths. The standard deviation of each one of the images in the selected
region of the continuum is shown in Fig. 6.
 |
Figure 6:
Standard deviations measured in the continuum around the Fe I
line at 5576 Å in the Stokes signals from a series of images of the disk
center.
The 3% of Stokes I is due to residual solar surface granulation (the
telescope was randomly displaced during the exposures). The values in the
other Stokes parameters are an indication of the photon noise and the errors
accumulated in the data reduction process |
In Stokes I the obtained value of around 3% is due to residual solar
granulation(the telescope was randomly displaced during the exposures).
Excepting this, the other 3 Stokes parameters show a common standard deviation
for the whole series of images of approximately 3 10-3 of the
intensity of the continuum. This noise
is slightly higher than the measured photon noise. For an exposure time
of 300 ms,
this number gives the polarimetric sensitivity of the instrument at the moment
of the observations (August 1998).
The previous measurement was done in the continuum in the disk center and
near a line not sensitive to the magnetic field. We also wanted to obtain
a value for the noise in a magnetically sensitive line in the presence
of field. We used the PCA techniques for this purpose (Rees et al. 1999)
and applied them to the series of 74 images taken on August 22nd for the
spectral domain around the 6301 Å Fe I line. Some 14000 profiles for
each one of the Stokes parameters were extracted from this series and
used to create a database, from which PCA procedures derived 4 sets of
eigenprofiles, one for each Stokes parameter. We also obtained the singular
values associated with each one of the eigenprofiles. These singular values
are an indication of the quantity of information contained in the
respective eigenprofiles. For instance, Fig. 7 shows the
ten first singular values obtained for Stokes Q. The entire set of eigenvalues
represents the total information contained in the whole set of Q profiles
extracted from
the referred data. In the figure, the eigenvalues have been normalized to the
sum of all the singular values, and its interpretation is very easy. For
example
the first eigenprofile contains on average 60% of the information present in
each profile
of the set taken into consideration. In fact it can be shown that this
first eigenprofile is the mean profile of the set. The second
eigenprofile will account for 17%, the third for 7% and so on.
In total, 97% of the information contained in any particular profile can
be reproduced by an adequate combination of the 10 first eigenprofiles.
 |
Figure 7:
The ten first singular values of the set of Stokes Q profiles for the
spectral region around 6301 Å from the observations of August 22nd. They
have been normalized to the sum of the whole set of eigenvalues. As explained
in the text, the value represents the amount of information contained on the
respective eigenprofile. The continuous line is for clarity purposes only |
The form of the curve in Fig. 7 gives us also an indication
of what kind of information each eigenprofile is taking into account.
PCA does not make the difference between signal and noise, but it concentrates
the organized patterns in the data in a few singular values: the bigger ones.
To reproduce the non-organized part of the data (i.e. noise and singular
physical configurations) PCA
require a lot of eigenprofiles. The respective singular values (the ones
with higher indices) will therefore
have similar values, indicating similar amounts of information. The
curve with the singular values becomes flattened when the respective
eigenprofiles are reproducing noise and not signal. In that sense PCA behaves
similarly to a Fourier transform, where the noise is concentrated into the
high frequencies, but it is much more efficient in distinguishing it. A more
detailed account of the properties of this mathematical technique can be
found in the book Numerical Recipes (Press et al. 1988), or in the papers by
Hansen (1992), Hansen et al. (1992) and
Hansen & Prost O'Leary (1993). For an application of this
technique to solar and stellar spectral profiles, see Rees et al. (1999).
Back to our noise measurements, ten eigenprofiles are more than enough to
reproduce the signal in the data discarding noise. However, to test the noise,
we have reconstructed
every profile present in the referred set with
a combination of up to 25
eigenprofiles, and calculated what was the residual left.
The results are presented in Figs. 8 and 9.
 |
Figure 8:
Standard deviation (in % of the continuum signal) of the differences
between the whole set of observed profiles and its reconstruction
using a variable number of eigenprofiles. The continuous line is for clarity
purposes only. The curves flatten from the moment they are reproducing noise.
No more than 10 eigenprofiles suffice to take into consideration most of
the information contained in the profiles |
 |
Figure 9:
Means (in % of the continuum signal) of the differences
between the whole set of observed profiles and its reconstruction
using a variable number of eigenprofiles. The continuous line is for clarity
purposes only. The means go to zero when the residuals are just white noise.
Once more (see previous figure) we see that at most 10 eigenprofiles are
necessary to reproduce the information contained in the whole set of profiles |
 |
Figure 10:
Image of the continuum for NOAA-8307 on August 22th (see the text).
The image is
pixels, each one spanning 0.4
 |
Once more we see in these figures that the first eigenprofile suffices to
reproduce each one of the observed profiles within 1 to 3% of the
signal of the continuum. This is however not enough and between 5 and 10
eigenprofiles must be used to flatten the four standard deviation curves
and drive the mean error to 0 (indicating a well centered noise). The error
committed in the reconstruction without noise is of 4 10-3
approximately, except for Stokes I which is kept at levels of 8 10-3 for such a small number of eigenprofiles. These
two numbers give the polarimetric noise level in the lines in the
presence of signal.
 |
Figure 11:
The continuum image at left indicates the slit position for which the
four Stokes parameters are shown (I,Q,U and V from left to right) for
the spectral region around 6301 Å |
 |
Figure 12:
The same as Fig. 11 except for the spectral region around
6149 Å |
The profiles considered in the series of data include all the polarization
signals of the observed
spot for two magnetic-sensitive lines and two telluric lines. The obtained
noise values are slightly higher (except for intensity for which it is
clearly bigger) than those measured in the
continuum. This is not surprising, as there is less signal in the lines than
in the continuum, and furthermore, the data comes from a scan over a
sunspot (while the continuum measurements were done for data taken on
the disk center and in the quiet sun) so the photon noise is also higher. Hence
the obtained result is consistent with the previous measurements.
Table 3:
Noise in data
Photon Noise |
2.4 10-3 |
Polarimetric sensitivity in the continuum |
3 10-3 |
Polarimetric sensitivity in the lines |
3-4 10-3 |
The three noise measurements, which we have summarized in Table 3,
are consitent with each other, and we can conclude that the photon noise level
was under 3 10-3 and that a polarimetric sensitivity of 3 to
4 10-3 of the intensity of the continuum was available at the moment of
these observations. It is very important to note that
these measurements do not take into account systematic errors.
Among these systematic errors we indicate the following:
- The observed difference between the equivalent widths for the two
paths, probably due to the presence of polarized scattered light in the
spectrograph;
- The errors in the correction of the spatial and temporal gains per pixel;
- The errors in the correction of the geometry between the two cameras;
- And the errors in the polarization analyser, which
will induce crosstalk among the Stokes parameters.
The first one is the most conspicuous in the reduced data and must
be the object of a thorough test for subsequent observations. The last one
is undoubtedly present, but the actual data does not allow a calibration nor
even an estimation of this effect. Without such a calibration
the expected residual instrumental polarization due to stresses in the
entrance window may not be removed.
 |
Figure 13:
The 4 Stokes parameters for a point of the first slit image in Fig. 11
at the spectral region around 6301 Å. Note the small inversion in the
V profile center, probably due to anomalous dispersion. Note also the
asymmetries in the Stokes profiles; they should be compared with those in the profiles
of the same slit position shown in Fig. 14 |
 |
Figure 14:
Asymetries in the profiles are evident at this point of the slit
extracted from the first image in Fig. 12 in the spectral region around
6149 Å. To be compared with the ones found in Figs. 13 and 15 |
 |
Figure 15:
Another set of Stokes profiles for the spectral region near
6149 Å extracted from the same slit as previous figure. Note the inverted
asymetries relative to that previous figure |
Up: First results from THEMIS mode
Copyright The European Southern Observatory (ESO)