The laboratory tests have been carried out at the ONERA (Office National d'Études et de
Recherche Aérospatiales) headquarter. A special device warms up the air confined in
a tank and generates the equivalent of a single-layer turbulent atmosphere. Temperature
and speed are calibrated to provide an adjustable set of values (D/r0,
), where D is the pupil diameter, r0 is the Fried parameter and
is the average wind speed. The disturbed wavefront originating from an
artificial light source (a binary star for example) is then restored with the BOA
adaptive optics (AO) system (Conan et al. 1998). The 88 actuators of the deformable
mirror allow seeing compensation at visible wavelength with a fast time response of
1 ms. The coronagraphic camera was installed downstream the AO bench.
Some of the limiting factors for the Lyot coronagraph (Lyot 1939) come from
inaccurate adjustment of the two main instrument components: the Lyot mask and the Lyot
stop. Setting up is made even harder when working on faint objects. In order to
understand some misadjustement effects such as alignment defects or longitudinal
misplacement of the coronograph's elements, one of us (L.A.) has developed a numerical
model of the optical set-up including not only the coronagraph itself, but also the
partial correction of the AO and the detector. As explained before, the Lyot stop tends
to concentrate some of the remaining light into a central peak in the coronagraphic
plane. Simulations have shown that a misalignment error for the stop exceeding of its diameter results in a nearly total extinction (
)
of the central peak. Some
experimental images also provided evidence of alignment drift. Further studies should
allow a more accurate testing of such effects, for providing a reproducible setup of
the instrument.
The coronagraphic frames are imaged on the detector with a high magnification (f/976) to achieve a fine sampling of 153 pixels per speckle area.
A Wynne device (Wynne 1979) has been included in the coronagraph to compensate the speckle chromatism as needed for dark-speckle observations with enough spectral bandwidth. The Wynne corrector, previously tested in a telescope run (Boccaletti et al. 1998b), provides a quasi-achromatic Airy pattern on a wide spectral band (650-850 nm).
The Lyot mask diameter is
(about 3 Airy radii) and the angular distance
between the primary and the companion is
.
The camera allows single photo-event detection and a very low dark count: less than 10
photons per 20 ms exposure, which corresponds to approximately 5 10-4photon/s/pixel for m pixels at
C.
Even though the centroiding electronics was limited to a reliable limit of 50000 photons/s (1000 photons per 20 ms short exposure), the low flux we were dealing with, allowed us to use the real-time facilities of the system such as live integration and display. The data files, recorded with CP20+, contain space-time coordinates of photo-events.
Copyright The European Southern Observatory (ESO)